Chloride

CHLORIDE ION SELECTIVE ELECTRODE

 
Description Combination Chloride ISE with Intermediate Junction reference system.
Concentration Range 1.8 to 35,000 mg/L Cl-
Linear Range 18 to 3,500 mg/L Cl-
Slope -57 mV/decade +/-3 mV at 25 oC
Response Time <30 seconds for 90% of final value from 18 to 35,000 mg/L Cl-
pH Range 2 to 14   (Note: An excess of OH- of 80 times the Cl- concentration produces a 10% error)
Main Interferences OH-, NH4+, S2O32-, Br -, I-, CN-, S2-
Temperature Range -0 to 60 oC (32 to 140 oF)
Sensor materials Polycrystalline solid state membrane
Body and sleeve Polypropylene/PEEK
Reference System Intermediate Junction (IJ) reference system is a double junction design. Ag/AgCl reference.
Dimensions 150 x 12 mm