Bromide

BROMIDE ION SELECTIVE ELECTRODE

 
Description Combination Bromide ISE with Intermediate Junction reference system.
Concentration Range 0.4 to 79,000 mg/L Br -
Linear Range 1 to 7,900 mg/L Br -
Slope -57 mV/decade +/-3 mV at 25 oC
Response Time <30 seconds for 90% of final value from 4.0 to 79,000 mg/L Br -
pH Range 2 to 14   (Note: An excess of OH - of 80 times the Br- concentration produces a 10% error)
Main Interferences OH-, Cl-, I-, CN-, S2-
Temperature Range -0 to 60 oC (32 to 140 oF)
Sensor materials Polycrystalline solid state membrane
Body and sleeve Polypropylene/PEEK
Reference System Intermediate Junction (IJ) reference system is a double junction design. Ag/AgCl reference.
Dimensions 150 x 12 mm
Ordering Information

Bromide ISE Analysis Kit

Part No
Bromide ISE with 1m cable and BNC plug
(go to the Connectors page for a listing of alternative connectors) Part No
Includes: 1 x 1000 mg/L Br ??? Standard (200mL) 121604
  1 x Bromide ISAB Solution (200mL) 121836
  1 x External Reference Electrolyte Solution 1M KNO345mL 121810
  1 x Bromide ISE Instruction Manual  
121600

 

Download this Specification Sheet in PDF format (0.34MB, .PDF)