CYANIDE ION SELECTIVE ELECTRODE

CYANIDE ION SELECTIVE ELECTRODE

Combination Cyanide ISE with Intermediate Junction reference system.

Specifications

Concentration Range 0.013 to 25 mg/L CN-

Linear Range 1 to 25 mg/L CN-

Slope -57 mV/decade +/-3 mV at 25 oC

Response Time <30 seconds for 90% of final value from 0.13 to 25 mg/L CN-

pH Range 9 to 14

Main Interferences I-, Br -, Cl-, OH-, S2-

Temperature Range -0 to 60 oC (32 to 140 oF)

Sensor materials Polycrystalline solid state membrane

Body and sleeve Polypropylene/PEEK

Reference System Intermediate Junction (IJ) reference system is a double junction design. Ag/AgCl reference.

Dimensions 150 x 12 mm