SMARTCHEM-ION3

Smartchem-ion3 Three Channel Specific Ion - Ph-mv-temperature Analyser

SMARTCHEM-ION3

The smartCHEM-Ion3 is a research grade Specific Ion/pH/mV meter with all the features expected in today's modern laboratories. It features 3 individual channels which can each be programmed for Specific Ions, pH or mV. 

Datalogging
The smartCHEM-Ion3 comes with a 1200 reading memory. Readings can be recorded manually, or automatically logged at preset intervals.

RS232 Serial Port Interface
The capabilities of the smartCHEM-Ion3 have been expanded with the RS232 port, which is standard equipment (cable sold separately). All readings stored in memory can be downloaded to an RS232 printer or PC. The RS232 port also allows the smartCHEM-Ion3 to log directly to the printer or PC. WinTPS Communication software for Windows 3.1 or 95, 98 & ME is available.

Automatic Stability Function (ASF)
ASF adds an extra level of versatility to the smartCHEM-Ion3. When ASF is activated, the smartCHEM-Ion3 monitors all parameters that are currently in use. When ALL parameters become stable, the readings are frozen on the display. ASF is automatically activated with extra fine settings during calibration, to ensure the most accurate possible results for the user.

ASF can also be used during Manual and Automatic Datalogging. When used during Automatic Datalogging, the smartCHEM-Ion3 will log until a stable end-point is reached for all parameters, then stop logging.

Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last calibration are stored in memory, along with the instrument’s serial number. This data can be displayed or sent to the RS232 port. All readings stored in memory are stamped with the date and time. Warning of failed calibration is provided.

A calibration alarm can be programmed to whichever parameters the user chooses. The alarm can be set for Daily, Weekly, 2 Weekly or 4 Weekly. The calibration alarm can also be switched off.

PIN Code Access
The Access Code is system is provided for those users who need to ensure the integrity of recorded data, calibration settings and so forth. The Access Code function will prevent anyone who does not know the code from entering the menu system. Keys that are available during normal measurement, (Store, Print, Autolog and F1 to F5 ) are still available without the Access Code. This means that others are still able to carry out day-to-day work.

Easy to use
The smartCHEM-Ion3 has a 240 x 64 dot graphic display. It features a user-friendly menu system, making the manual virtually unnecessary. Full-text help and error messages are provided. All readings and messages are provided in large, bold characters. 

Automatic Calibration
Where necessary, the user is able to define the standards to be used for calibration. For pH calibration, the smartCHEM-Ion3 is programmed to automatically recognise the most popular buffers, making calibration a breeze. Any non-standard buffer value can be entered during calibration. 

Karl Fischer Measurements
A polarisation output is provided for Karl Fischer moisture titrations. 

Reliability and Service
The smartCHEM-Ion3 is fully waterproof to IP65, so it is suitable for even the harshest laboratory or factory environments. 
TPS has been manufacturing top quality, reliable instruments and electrodes since 1968. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Our service is also of the highest standard. Spares and accessories are always held in stock, and our modern service facilities ensure a rapid turn-around for maintenance and repair work.

Product price

Product ID Product options Price
121164 smartCHEM-Ion3 Triple Channel Ion-pH-mV-Temp Meter *Ion Sensor not included* $1,273.00

Specifications

Specific Ions
Ranges : Auto-ranging in units of ppM, ppK, % and Exponential Notation
Resolution : User selectable for 3 significant digits, 2 significant digits or Auto-rounding
Accuracy : +/-Least significant digit

pH
Ranges : 
0.000 to 14.000 pH
0.00 to 14.00 pH
Resolution : 
0.001 pH
0.01 pH
Accuracy : 
+/-0.002 pH
+/-0.01 pH

Millivolts & Relative Millivolts
Range : 0 to +/-400.0 and 0 to +/-1500 mV (auto-ranging)
Resolution : 
0.1 mV
1 mV
Accuracy : 
+/-0.15 mV
+/-1 mV

Temperature Range : -10.0 to 120.0 oC
Resolution : 0.1 oC
Accuracy : +/-0.2 oC

Temperature Compensation
0 to 100 oC, automatic or manual.

Calibration
Automatic calibration for all parameters.
User can define standards where applicable.

Memory
2730 readings including date and time.

Automatic Datalogging
User-set for one reading every 2 to 90 seconds, minutes or hours. 

RS232 Serial Port Output
1200, 9600, 19200 and 38400 Baud.
8 Bits, No Parity, 1 Stop Bit, XON/XOFF Protocol

Recorder Output (optional)
Max 0 to 2000mV output for any one parameter at one time. Output impedance approx 1000 Ohms.
Resolution approx 2 mV.

Clock
24 Hour Calendar clock.
Displays date, month, year, hours, minutes, seconds.

Good Laboratory Practices
Date, time and value of last successful calibration for each parameter are stored. This information can be recalled or sent to the RS232 port at any time.

Calibration Alarm
Calibration Alarm can be activated for parameter of user's choice. Calibration Alarm can be set to Daily, Weekly, 2 Weekly or 4 Weekly. 

Display
240 x 64 dot graphic LCD with large, bold digits, user-friendly menu system and full-text help and error messages. 

Enclosure
Tough, high impact ABS. Splash resistant one-piece lid.

Dimensions
240 x 180 x 105 mm

Mass
Instrument only : Approx 0.8 kg
Full Kit : Approx 3.0 kg

Environment
Temperature : 5 to 45 oC
Humidity : 0 to 90% R.H.

Power
12V DC via AC/DC adaptor.
AC/DC adaptor to suit specified country is supplied.

Ordering Information

Part No
smartCHEM-Ion Ion/pH/mV-Temp....121163

Kit Includes
(Ion sensors sold separately.)
Temperature/ATC Sensor..........121248
AC/DC Power Adaptor.............130037
Manual..........................130050

Optional Sensors
Click the following links for details on the TPS range of sensors...
Specific Ion and Reference sensors
pH Sensors
Redox (ORP) Sensors

Part No
Options and Accessories
Computer Interface:
Serial cable for computer.......130041
Serial to USB adaptor cable.....130087
WinTPS Software for Windows.....130086

Recorder:
Recorder Output & cable.........130028

Sensor Holder:
Flexible Sensor Holder..........130088

Download this Specification Sheet in PDF format (0.37MB, .PDF)
Download the manual for this instrument (1.61MB, .PDF)

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.