PROCHEM-D

Prochem-d Dissolved Oxygen Controller With Dual Set Points

PROCHEM-D

 

The NEW proCHEM Series controllers provide an economical answer to most process monitoring, control or recording requirements.
Dual Set Points are provided. The default setup for these are as "Too High" and "Too Low" control points. The user can also re-set them so that they are both "Too High" or both "Too Low" control points. An example of this type of setup is where the proCHEM unit has one point set to control the process, and a second point beyond that set to activate an alarm.

Separate LED's indicate when either of the Set Points has been exceeded. The output relays are rated to 240V and are fused at 1 Amps. Higher currents can be switched with an external contactor. All proCHEM units are easy to install. They are housed in a rugged plastic case which is non corrosive, waterproof and dustproof to IP65. All controls can be accessed via the hinged, clear cover. The large LCD display is easy to read, even from a distance. 

The proCHEM series is designed for maximum flexibility. 4 to 20 mA and 0 to 1 V DC outputs are provided as standard.

For a total solution to your control requirements, TPS also offers a peristaltic pump, the easyFLOW-VS.

The proCHEM series is proudly designed and manufactured in Australia. The TPS Quality System has been certified in accordance with AS/NZS ISO9001 standards. Since 1968, TPS has built a reputation for excellent backup service that is second to none.

Product price

Product ID Product options Price
113122 proCHEM-DO2 Controller, 240V AC In, Dual Limits, 4-20mA Out $719.00
113122/5 proCHEM-DO2 with 5 metre cable & YSI DO2 sensor $1,604.00

Specifications

Ranges
0 to 199.9 % Saturation
OR 0 to 19.99 ppM (mg/L) (Factory-set)

Resolution
0.1 % Saturation
OR 0.01 ppM (mg/L)

Accuracy
+/-0.2% of full scale

Linearity
+/-0.1%

Repeatability
+/-0.2%

Ambient Drift
<0.05% span/oC

Long Term Drift
<0.2% per year

Temperature Compensation
Dual automatic temperature 
compensation system,(sensor limit 45 oC).

Zero Range
+/-10%

Span Range
70 to 130%

Sensor
YSI5739 Polarographic sensor with inbuilt ATC sensor.

Enclosure
Polycarbonate, waterproof to IP65

Display
12.5mm LCD

Outputs
4 to 20 mA (500 Ohms max load)
0 to 1 V DC (0 to 10 V DC optional)
Contacts : One optional relay with changeover contacts. Rated to 240V, 1 Amp

Isolation
Galvanic isolation of sensor input
Current and voltage outputs are not isolated 

Power
220/240 V AC, 50/60Hz
(110/120 V AC available)
12V DC power optional.

Dimensions
172 x 160 x 71mm

Mass
Instrument only : Approx 800g 
Full Kit : Approx 1.5kg

Instrument Operating Environment
Temperature : 0 to 45 OC
Humidity : 0 to 95% R.H.

Ordering Information

Part No
5 metre bundle 113122/5:
proCHEM-DO2 5m.................................113122/5

Kit Includes:
DO2 Sensor YSI5739 (no cable)..................123204
5m cable for Sensor (YSI5739)..................123210
Membrane & filling solution kit................123300
Manual.........................................130050

Options and Accessories
Dissolved Oxygen Sensor Maintenance:
Membrane & filling solution kit................123300
Filling Solution only, 45mL....................123303
Zero calibration Sodium Sulphite...............123302
Rejuvenation kit...............................123037
Diaphragm replacement kit......................123304
(123304 for YSI5739 Field Sensor Only) 

Cable Extension
Extended cable/metre...........................130040

Instrument Options
12V DC Power option............................130072

Download this Specification Sheet in PDF format (0.42MB, .PDF)
Download the manual for this instrument (0.33MB, .PDF)

Hint : Right click on a link, then select "Save Target As..." or "Save Link As..." to save the file to your own computer.